  • 200VA AC test capacity
  • 240×64 ice blue dot matrix LCD
  • Manual / automatic mode
  • Function button for quick selection
  • High-intensity flash for warning and status indication
  • Safety INTERLOCK function
  • Zero Crossing Turn-on Operation
  • Adjustable lead time
  • True RMS current measurement
  • High resolution: 1uA for measuring current, 2V for setting voltage
  • PWM amplifier to improve energy efficiency and reliable testing
  • Max. 100 memory block for test condition (step) setting. And each step can be named separately.
  • External terminal on front panel for “start” and “stop” operation
  • Interface: RS-232C, USB device, signal I/O and GPIB (optional)
Category:

A solid foundation for reliable, safe and practical safety testing GW Instek launches a new series of safety testers (hi-pot tester / hipot tester) with high durability and multiple user protection design. The GPT-9800 series meets various safety standards, such as IEC, EN, UL, CSA, GB, JIS and other safety-related requirements for the safety testing of electronic products and components. The GPT-9801 can pass AC tests. The series of safety testers is built on a highly efficient PWM amplifier platform with a maximum output capacity of AC 200VA to impede the influence of voltage fluctuations from the input AC power source. This ensures a stable voltage supply for all GPT-9800 series tests. Focusing on user protection, the GPT-9800 series uses a combination of hardware and software controls to enhance safety: a self-check is performed to ensure that all functions and operations are under normal conditions each time the tester is switched on; the Zero Crossing Turn-On design imposes the output voltage to always start at the zero crossing of a sine wave to avoid the impact of over-voltage output; and the output voltage is automatically switched off (within 150μs) when abnormal output voltages are detected or when upper or lower current limits are reached during testing. To protect the operator from dangerous injury, the GPT-9800 series automatically discharges the DUT after testing (within 200ms) to eliminate excessive voltage on the DUT. To further ensure safety, the interlock connector can also be used as dual protection to prevent unintended operation. Equipped with a simple and clear panel layout, a high-resolution dot-matrix LCD display and colour LED indicators, the GPT-9800 series allows operators to interpret measurement results easily and quickly. All major test functions, including AC resistance (AC 5kV / 40mA), DC resistance (DC 6kV / 10mA), insulation resistance (DC 50V ~ 1000V) and earth connection (AC 30A max.), Are performed under high stability voltage output with high-resolution measurement results. Furthermore, the test duration, lead time and upper/lower limits of the output voltage are fully adjustable to accommodate a wide range of safety tests with accurate measurement results. Other functions and features of the GPT-9800 include: open circuit detection for earth connection tests to get the correct measurement result with confidence; 100 sets of memory to store panel settings, which can be recalled for individual tests individually or for automatic tests in sequence; an external output on/off connector in the front panel and a signal I/O port in the rear panel provided as a means for remote start/stop control of the safety tester; RS-232C, USB and GPIB interfaces (optional) available for PC remote control and recording of test results.

 
GPT-9800 Specifications

The specifications apply when the GPT-9800 is powered on for at least 30 minutes under +20°C~+30°C.

 

Model Func. AC WITHSTANDING DC WITHSTANDING INSULATION RESISITANCE GROUND BOND
GPT-9804 P P P P
GPT-9803 P P P O
GPT-9802 P P O O
GPT-9801 P O O O

 

AC WITHSTANDING

 

Output-Voltage Range 0.050kV~ 5,000kV ac
Output-Voltage Resolution 2V/step
Output-Voltage Accuracy ± (1% or setting +5V) [no load].
Maximum Rated Load 200 VA (5kV/40mA)
Maximum Rated Current 40mA (0.5kV < V ≦ 5kV)

10mA (0.05kV ≦ V≦ 0.5kV)

Output-Voltage Waveform Sine wave
Output-Voltage Frequency 50 Hz / 60 Hz selectable
Voltage regulation ± (1% or reading +5V) [full load → no load].
Voltmeter Accuracy ± (1% or reading+ 5V)
Current Measurement Range 0.001mA~040.0mA
Current Best Resolution 0.001mA / 0.01mA / 0.1mA
AC Current Measurement Accuracy ±(1.5% of reading +30 counts) when HISET < 1.11mA

±(1.5% of reading +3 counts) when HISET ≧ 1.11mA

Window Comparator Method Yes
ARC Detect Yes
RAMP (Ramp-Up Time) 0.1~999.9s
TIMER (Test Time)* OFF, 0.5s~999.9s
GND ON/OFF


DC WITHSTANDING

 

Output-Voltage Range 0.050kV~ 6,000kV dc
Output-Voltage Resolution 2V/step
Output-Voltage Accuracy ± (1% or setting +5V) [no load].
Maximum Rated Load 50W (5kV/10mA)
Maximum Rated Current 10mA (0.5kV < V ≦ 6kV)

2mA (0.05kV ≦ V≦ 0.5kV)

Voltage regulation ± (1% or reading +5V) [full load → no load].
Voltmeter Accuracy ± (1% or reading+ 5V)
Current Measurement Range 0.001mA~010.0mA
Current Best Resolution 0.001mA / 0.01mA / 0.1mA
DC Current Measurement Accuracy ±(1.5% of reading +30 counts) when HISET < 1.11mA

±(1.5% of reading +3 counts) when HISET ≧ 1.11mA

Window Comparator Method Yes
ARC Detect Yes
RAMP (Ramp-Up Time) 0.1~999.9s
TIMER (Test Time)* OFF, 0.5s~999.9s
GND ON/OFF


INSULATION RESISITANCE

 

Output-Voltage 50V~1000V dc
Output-Voltage Resolution 50V/step
Output-Voltage Accuracy ± (1% or setting +5V) [no load].
Resistance Measurement Range 1 Mohm~ 9500 Mohm
Test voltage Measurable Range Accuracy
50V≦V≦450V 1 Mohm~ 50 Mohm

51 Mohm~ 2000 Mohm

±(5% of reading +1 count)

±(10% or reading +1 counts)

500V≦V≦1000V 1M ohm~ 500 Mohm

501 Mohm~ 9500 Mohm

±(5% of reading +1 count)

±(10% or reading +1 count)

Window Comparator Method Yes
Output Impedance 600 kohm
RAMP (Ramp-Up Time) 0.1~999.9s
TIMER (Test Time) 1s~999.9s
GND OFF (fixed)


GROUND BOND

 

Output-Current Range 03.00A~30.00A ac
Output-Current Resolution 0.01A
Output-Current Accuracy 3A≦I≦8A : ± (1% or setting +0.2A)

8A<I≦30A : ± (1% or setting +0.05A)

Test-Voltage 6Vac max (open circuit)
Test-Voltage Frequency 50 Hz / 60 Hz selectable
Resistance Measurement Range 10mW~650.0mW
Resistance Measurement Resolution 0.1mW
Resistance Measurement Accuracy ± (1% or reading +2mW)
Window Comparator Method Yes
TIMER (Test Time) 0.5s~999.9s
GND OFF (fixed)
Test method Four Terminal

MEMORY

Single Step Memory MANU : 100 blocks
Automatic Testing Memory AUTO : 100 blocks, menu per car : 16

 INTERFACE

RS232 Standard
USB (Device) Standard
GPIB Option
Remote Terminal (Front) Standard
SIGNAL IO Standard

 General

DISPLY 240 x 64 dot matrix LED backlight LCD
POWER SOURCE AC100V/120V/220V/230V ±10%, 50/60Hz
POWER CONSUMPTION 500VA max.
DIMENSIONS & WEIGHT 330(W) x 148(H) x 452(D) mm,

approx.19kg(Max)