FLUKE i6000sFlex AC current measuring probe

Producer: Fluke

The i6000sflex is an AC current measurement probe that operates on the Rogowski principle. It can be used for current measurements up to 6000 A when used in combination with oscilloscopes, recorders or data loggers. Thanks to its flexible lightweight measuring head, it is quick and easy to install in hard-to-reach places.

Electrical characteristics
Measuring ranges current
   60 A / 600 A / 6000 A ACRMS
Output sensitivity (AC-coupled)
   50 mV / 5 mV / 0.5 mV per A
Load impedance
   100 kOhm minimum
Accuracy (at +25 °C)
   ±1% of the range (45 – 65 Hz)
Linearity (from 10% to 100% of range)
   ±0.2% of measured value
Noise
   8 mV ACRMS (60 A)
   2 millivolts ACRMS (600/6000A)
Frequency range
   10 Hz to 50 kHz (-3 dB)
Phase error
   < ±1° (45 – 65Hz),
   ±10° (at 20 kHz)
Positional sensitivity (with cable > 25 mm from the coupling)
   ±2% of the range
External field (with cable > 200 mm from the sensor head)
   ±1% of the range
Nutrition
   2 x AA MN 1500 LR6 alkaline battery
   400 hours, indication of low battery voltage or special external power supply
Temperature coefficient
   ±0.08% of measured value per C
Operating voltage (see under Safety standards)
   600 V ACRMS or DC
General characteristics
Length of measuring head cable (double insulated)
   610 mm (24”) I6000s Flex-24
915 mm (36”) I6000s Flex-36
Diameter of the head cable
   14.3 mm
Bending radius measuring head
   38.1 mm
Cable length (measurement head to electronics)
   2 metres
Output connector
   0.5m cable with
BNC safety connector
Including adapter for 4mm
Safety connector
Operating temperature range
   -20 to +90 °C (measuring head)
-20 to +85 °C (electronics)
Storage temperature range
   -40 to +105 °C (measuring head)
-20 to +85 °C (electronics)
Relative humidity during operation
   15% to 85% (non-condensing)
Weight
   180 g (cup), 190 g (electronics)
Safety standards
BS EN 61010
   1: 2001
BS EN 61010
   2-032: 2002
BS EN 61010
   031: 2002
600 VRMS, category III, pollution degree 2
   
Comment
   The use of the probe on non-isolated conductors is restricted to 600 V AC RMS or DC and frequencies below 1 kHz.
EMC standards
BS EN 61326:  1998 +A1, A2 and A3
Download datasheet

Need advice?

Contact our product specialist Thiara for help in choosing the right instrument.
Contact Thiara

Select at least 2 products